Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1001920120510020120
Journal of Korean Neurosurgical Society
2012 Volume.51 No. 2 p.120 ~ p.123
Preliminary Report of Three-Dimensional Reconstructive Intraoperative C-Arm in Percutaneous Vertebroplasty
Shin Jae-Hyuk

Jeong Je-Hoon
Abstract
Objective: Percutaneous vertebroplasty (PVP) is usually carried out under three-dimensional (2D) fluoroscopic guidance. However, operative complications or bone cement distribution might be difficult to assess on the basis of only 2D radiographic projection images. We evaluated the feasibility of performing an intraoperative and postoperative examination in patients undergoing PVP by using three-dimensional (3D) reconstructive C-arm.

Methods: Standard PVP procedures were performed on 14 consecutive patients by using a Siremobil Iso-C3D and a multidetector computed tomography machine. Post-processing of acquired volumetric datasets included multiplanar reconstruction (MPR) and surface shaded display (SSD). We analyzed intraoperative and immediate postoperative evaluation of the needle trajectory and bone cement distribution.

Results: The male : female ratio was 2 : 12; mean age of patients, 70 (range, 77-54) years; and mean T score, -3.4. The mean operation time was 52.14 min, but the time required to perform and post-process the rotational acquisitions was 7.76 min. The detection of bone cement distribution and leakage after PVP by using MPR and SSD was possible in all patients. However, detection of the safe trajectory for needle insertion was not possible.

Conclusion: 3D rotational image acquisition can enable intra- or post-procedural assessment of vertebroplasty procedures for the detection of bone cement distribution and leakage. However, it is difficult to assess the safe trajectory for needle insertion.
KEYWORD
Vertebroplasty, Intraoperative 3D imaging, Complication, Feasibility
FullTexts / Linksout information
  
Listed journal information
SCI(E) ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø