Journal Of The Korea Institute Of Oriental Medical Diagnostics
2005 Volume.9 No. 1 p.131 ~ p.135
The Problems in the Measurement of DC Potential on Meridian Skin Area
Heo Ik-Boom

Lee W.C.
Yin Chang-Shik
Lee Yoo-Jeong
Abstract
Background and purpose: The measurement of direct current (DC) potential on skin area of meridian has recently been adopted to explore the electrophysiological characteristics of meridian system, But there exists two problems to be explained; the entity of the DC potential to be measured and the characteristics of electrode gel interface of measurement system, It is not clear whether the DC potential reflect, at least hypothetically, the entity of meridian, and if there exist any unstable factor in the DC potential measurement system.

Methods: In this study, we designed an electronic circuit model of skin and applied known DC potential sources to the electrode interface of the skin model.

Results: The result showed that the measured DC potential changed according to the time, and the same phenomenon was observed when the electrode gel was replaced with an electric condenser. It is suggested that the measurement of DC potential on electrode gel interface is very difficult and produces unstable values due to the capacity effect of electrode gel.

Conclusion: Further studies on the DC potential evaluation in the context of meridian study should consider and bypass this problem.
KEYWORD
Meridian, DC potential, Electrode, Conduction Gel