Study on the Repeatability and Reproductivity of a Moire Body Shape Analyser
Lee Dong-Yup

Park Young-Bae
Oh Hwan-Sup
Abstract
Background : As each human has a look in the face of oneself, he or she has a look of him or herself in the shape of the body also. And for the shape of the body which gives a big clue in diagnosis in musculoskeletal disorders. Therefor many means are used and developed for diagnosis through body shape or posture analysis for musculoskeletal disorders. X-ray, CT, MRI has been used for diagnosis through image in this way to tell about the inside of the human body. On the other side, moire topography was used for information about the exterior of the human body, but yet only minimal information such as the number of contour lines in each side was available. Therefor there were a few studies to use moire topography or other methods to get information about the surface of the human body in numeric values. The instrument used in this study which is a laser projection moire, is another trial to get numeric data about the surface of the human body. The instrument is composed of laser projector and a computer software to recompose and analyse the image data into depth, height, angle and length.

Objectives : The study was focused on whether the instrument is reliable for clinic use, and to seek the proper environment and posture for the examination, and among the data the software provides, which items are more reliable and useful.

Methods : For reproductivity and repeatability, 4 testers tested 2 persons. And to how if the body shape changes according to the posture and which posture gives the most reliable data, the test was performed in 6 different positions.
Results : Result, the instrument showed sufficient repeatability and reproductivity for clinical use. And among the items the software provides, the length of the back, the angle of the back in the sagittal and coronal plane showed reliable results. And there was difference in the results according to the posture, and Therefor, in following studies using this instrument or similar type of posture analysing instruments, the length of the back, the angle of the back in the sagittal and coronal plane could be reliable item to use.
KEYWORD
moire, body shape analysis, IBS, scoliosis, spine, oriental medicine, repeatability
FullTexts / Linksout information
Listed journal information