Evaluation of methods for estimating the pulse reflection site with cardiovascular simulator
|
|
Lee Ju-Yeon
Shin Sang-Hoon
|
|
Abstract
|
|
|
Objective: Wave reflection is an important factor that determines the shape of the pulse wave. The purpose of this study is to compare the conventional method used for estimating the reflection site of pulse with a cardiovascular simulator.
Methods: In this study, cardiovascular simulator with one elastic tube was used. The pressure and flow was measured simultaneously at three different points. The measured data were used to the conventional methods to estimate the pulse wave reflection site. The results were compared with the known length which were the distances from the measured points to the end of tube.
Results & Conclusions: There is a significant error with the time domain method. While, the reflection site with the frequency domain method was similar to the actual reflection site.
|
|
KEYWORD
|
|
Pulse wave , Pulse reflection site , Cardiovascular simulator
|
|