KMID : 1004520100260040373
|
|
Journal of Dental Rehabilitation and Applied Science 2010 Volume.26 No. 4 p.373 ~ p.388
|
|
Fracture Analysis of Implant Components Using Scanning Electron Microscope. Part I - Implant Retaining Screw
|
|
Lim Kwang-Gil
Kim Dae-Gon Cho Lee-Ra Park Chan-Jin
|
|
Abstract
|
|
|
Fracture causes serious problems in many instance of prosthetic failures. But it is hard to find the definite causes when fractures occur. Fractography encompasses the examination of fracture surfaces that contain features resulting from the interaction of the advancing crack with the microstructure of the material and the stress fields. All fractured specimens(implant retaining screw) retrieved from Gangneung-Wonju national university dental hospital for 3 years(from 2007 to 2009). After pretreatment of samples, the scanning electon microscope were used for surface examination and fracture analysis. In case of most of the fractured specimens, fracture took place by fatigue fracture and fractured surface represents fatigue striation. Fatigue striation indicate the progression of the crack front under cyclic loading, are characteristic of stage 2 crack growth. The site of crack initiation and stage 1 crack growth were not easily identified in any of the failure, presumably because of the complex microstructural features of the polycrystalline sample. In case of fractured by overload, dimpled or cleavage surface were observed. Using the interpretation of characteristic markings(ratchet mark, fatigue striation, dimple, cleavage et al) in fracture surfaces, failure events containing the crack origin, crack propagation, material deficiency could be understand. Using the interpretation of characteristic markings in fracture surfaces, cause and mechanism of fractures could be analyzed.
|
|
KEYWORD
|
|
Fracture analysis, implant retaining screw, scanning electron microscope
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|