KMID : 1004520170330030189
|
|
Journal of Dental Rehabilitation and Applied Science 2017 Volume.33 No. 3 p.189 ~ p.198
|
|
Stress distribution of implants with external and internal connection design : a 3-D finite element analysis
|
|
Chung Hyun-Ju
Yang Sung-Pyo Park Jae-Ho Park Chan Shin Jin-Ho Yang Hong-So
|
|
Abstract
|
|
|
Purpose: This study aims to analyze the stress distribution of mandibular molar restoration supported by the implants with external hex and internal taper abutment connection design.
Materials and Methods: Models of external connection (EXHEX) and internal connection (INCON) implants, corresponding abutment/crowns, and screws were developed. Supporting edentulous mandibular bony structures were designed. All the components were assembled and a finite element analysis was performed to predict the magnitude and pattern of stresses generated by occlusal loading. A total of 120 N static force was applied both by axial (L1) and oblique (L2) direction.
Results: Peak von Mises stresses produced in the implants by L2 load produced 6 - 15 times greater than those by L1 load. The INCON model showed 2.2 times greater total amount of crown cusp deflection than the EXHEX model. Fastening screw in EXHEX model and upside margin of implant fixture in INCON model generated the peak von Mises stresses by oblique occlusal force. EXHEX model and INCON model showed the similar opening gap between abutment and fixture, but intimate sealing inside the contact interface was maintained in INCON model.
Conclusion: Oblique force produced grater magnitudes of deflection and stress than those by axial force. The maximum stress area at the implant was different between the INCON and EXHEX models.
|
|
KEYWORD
|
|
implant, abutment-implant connection, finite element analysis, stress distribution, biomechanics
|
|