KMID : 1011420160210030281
Journal of Korean Ophthalmic Optics Society
2016 Volume.21 No. 3 p.281 ~ p.288
The Analysis of Optical Quality in Central Serous Chorioretinopathy by OQAS
Choi Jong-Kil

Lee Kyung-Min
Kim Hee-Sung
Kim So-Ra
Park Mi-Jung
Abstract
Purpose : The present study was aimed to investigate the change of optical quality in fully cured eyes from central serous chorioretionopathy(CSC) and CSC under treatment, respectively.

Methods : A total of 136 eyes(68 subjects) was divided into CMC under treatment and its asymptomatic contralateral eye, fully cured eyes after being diagnosed with CMC and its contralateral eye, and normal eyes. Objective scattering index, cut-off value in modulation transfer function, focusing rate, PSF(Point Spread Function) width at 50%, 10% and the contrast sensitivity were measured and analyzed.

Results : CMC under treatment showed statistically significant decrease in all evaluations of optical quality compared with asymptomatic contralateral eyes, fully cured eyes and normal eyes. Although there was no significant difference in optical quality between asymptomatic contralateral eyes of CMC under treatment and fully cured eyes, there was tendence in optical quality decreasing compared with normal eyes.

Conclusion : From the study, it was revealed that the optical quality was different according to the treatment phase in CMC. The use of optical quality assessment for the decision of complete cure and the evaluation of prognosis in CMC is also proposed.
KEYWORD
Central serous chorioretinopathy, OQAS, OSI, MTF, Strehl ratio, Contrast