KMID : 1011820180590040275
Investigative and Clinical Urology
2018 Volume.59 No. 4 p.275 ~ p.279
Proof of concept: Exposing the myth of urethral atrophy after artificial urinary sphincter via assessment of circumferential recovery after capsulotomy and intraoperative pressure profiling of the pressure regulating balloon
Pearlman Amy Marcia

Rasper Alison Marie
Terlecki Ryan Patrick
Abstract
Purpose: Rate of continence after artificial urinary sphincter (AUS) placement appears to decline with time. After appropriate workup to exclude inadvertent device deactivation, development of urge or overflow incontinence, and fluid loss, many assume recurrent stress urinary incontinence (rSUI) to be secondary to nonmechanical failure, asserting urethral atrophy as the etiology. We aimed to characterize the extent of circumferential urethral recovery following capsulotomy and that of pressure regulating balloon (PRB) material fatigue in men undergoing AUS revision for rSUI.

Materials and Methods: Retrospective review of a single surgeon database was performed. Cases of AUS removal/replacement for rSUI involving ventral subcuff capsulotomy and intraoperative PRB pressure profile assessments were identified.

Results: The described operative approach involving capsulotomy was applied in 7 patients from November 2015 to September 2017. Mean patient age was 75 years. Mean time between AUS placement and revision was 103 months. Urethral circumference increased in all patients after capsulotomy (mean increase 1.1 cm; range 0.5?2.5 cm). Cuff size increased, remained the same, and decreased in 2, 3, and 2 patients, respectively. Six of 7 patients underwent PRB interrogation. Four of these 6 PRBs (66.7%) demonstrated pressures in a category below the reported range of the original manufacturer rating.

Conclusions: Despite visual appearance to suggest urethral atrophy, subcuff capsulotomy results in increased urethral circumference in all patients. Furthermore, intraoperative PRB profiling demonstrates material fatigue. Future multicenter efforts are warranted to determine if capsulotomy, with or without PRB replacement, may simplify surgical management of rSUI with reductions in cost and/or morbidity.
KEYWORD
Atrophy, Urinary incontinence, Urinary sphincter, artificial
FullTexts / Linksout information
Listed journal information
SCI(E) MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø