KMID : 1011920180190010033
|
|
International Journal of Arrhythmia 2018 Volume.19 No. 1 p.33 ~ p.35
|
|
Short-Long R-R Sequence Induced Implantable Cardioverter Defibrillator Shock by Dislodged Lead Even Having Normal Interrogation Parameter
|
|
Kim Ki-Hun
Seol Sang-Hoon Kim Seung-Hwan Kim Dong-Kie Kim Doo-Il
|
|
Abstract
|
|
|
Inappropriate shocks from an implantable cardioverter defibrillator (ICD) can cause potentially dangerous ventricular arrhythmias and impaired quality of life. We describe a case in which a dislodged lead caused inappropriate ICD shocks through simultaneous sensing of atrial and ventricular signals. Interestingly, repeated short-long R-R sequences were recorded, but ICD interrogation parameters were usually unchanged.
|
|
KEYWORD
|
|
Defibrillators, Implantable, Lead
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|