KMID : 1023420150150050659
Journal of Dental Hygiene Science
2015 Volume.15 No. 5 p.659 ~ p.665
Effects of Flushing in Dental Unit Waterlines on the Bacterial Contamination Level
Yoon Hye-Young

Lee Si-Young
Abstract
A wide variety of methods have been used to control Dental Unit Waterline (DUWL) contamination. Among the methods, flushing is mainly used because it is simple and easy to use. Generally, flushing of DUWL for 20 or 30 sec before using high speed handpieces or scalers is recommended. However, the appropriateness of flushing time was not investigated thoroughly. The purpose of this study was to check the effective time of flushing for decreasing bacterial contamination. Seven dental unit chairs were randomly selected in student clinical simulation laboratory for this experiment. DUWLs were continuously flushed and water samples were collected at an interval of 30 seconds for 15 minutes. From five dental unit chairs, water samples were collected every 10 seconds for 1 minute. Bacterial levels in water samples were examined by the culture method on R2A plates. After 10 second flushing of DUWLs, the number of bacteria significantly reduced and decreased continuously up to 40 seconds. However, even after the water was flushed for 15 minutes, the bacterial contamination level was not reduced below recommended bacteria level, 200 CFU/ml. In addition to flushing, the periodic chemical disinfection is required to control the DUWL water to the recommended level.
KEYWORD
Biofilm, Decontamination, Flushing, Water microbiology, Water quality