KMID : 1024020120420010025
Imaging Science in Dentistry
2012 Volume.42 No. 1 p.25 ~ p.33
Evaluation of accuracy of 3D reconstruction images using multi-detector CT and cone-beam CT
Kim Mi-Ja

Huh Kyung-Hoe
Yi Won-Jin
Heo Min-Suk
Lee Sam-Sun
Choi Soon-Chul
Abstract
Purpose: This study was performed to determine the accuracy of linear measurements on three-dimensional (3D) images using multi-detector computed tomography (MDCT) and cone-beam computed tomography (CBCT).

Materials and Methods: MDCT and CBCT were performed using 24 dry skulls. Twenty-one measurements were taken on the dry skulls using digital caliper. Both types of CT data were imported into OnDemand software and identification of landmarks on the 3D surface rendering images and calculation of linear measurements were performed. Reproducibility of the measurements was assessed using repeated measures ANOVA and ICC, and the measurements were statistically compared using a Student t-test.

Results: All assessments under the direct measurement and image-based measurements on the 3D CT surface rendering images using MDCT and CBCT showed no statistically difference under the ICC examination. The measurements showed no differences between the direct measurements of dry skull and the image-based measurements on the 3D CT surface rendering images (P>.05).

Conclusion: Three-dimensional reconstructed surface rendering images using MDCT and CBCT would be appropriate for 3D measurements.
KEYWORD
Reproducibility of Results, Three-Dimensional Image, Skull