Imaging Science in Dentistry
2013 Volume.43 No. 3 p.185 ~ p.190
Effect of digital noise reduction on the accuracy of endodontic file length determination
Mehdizadeh Mojdeh

Khademi Abbas Ali
Shokraneh Ali
Farhadi Nastaran
Abstract
Purpose: The aim of the present study was to evaluate the measurement accuracy of endodontic file length on periapical digital radiography after application of noise reduction digital enhancement. Materials and Methods: Thirty-five human single-rooted permanent teeth with canals measuring 20-24 mm in length were selected. ISO #08 endodontic files were placed in the root canals of the teeth. The file lengths were measured with a digital caliper as the standard value. Standard periapical digital images were obtained using the Digora digital radiographic system and a dental X-ray unit. In order to produce the enhanced images, the noise reduction option was applied. Two blinded radiologists measured the file lengths on the original and enhanced images. The measurements were compared by repeated measures ANOVA and the Bonferroni test (¥á=0.05). Results: Both the original and enhanced digital images provided significantly longer measurements compared with the standard value (P<0.05). There were no significant differences between the measurement accuracy of the original and enhanced images (P>0.05). Conclusion: Noise reduction digital enhancement did not influence the measurement accuracy of the length of the thin endodontic files on the digital periapical radiographs despite the fact that noise reduction could result in the elimination of fine details of the images.
KEYWORD
Radiography, Dental, Digital, Endodontics, Noise, Radiographic Image Enhancement
FullTexts / Linksout information
  
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed