KMID : 1024020190490040295
Imaging Science in Dentistry
2019 Volume.49 No. 4 p.295 ~ p.299
A clinical pilot study of jawbone mineral density measured by the newly developed dual-energy cone-beam computed tomography method compared to calibrated multislice computed tomography
Kim Hyun-Jeong

Kim Ji-Eun
Choo Ji-Yeon
Min Jeong-Hee
Chang Sung-Ho
Lee Sang-Chul
Pyun Woong-Beom
Seo Kwang-Suk
Karm Myong-Hwan
Koo Ki-Tae
Rhyu In-Chul
Myoung Hoon
Heo Min-Suk
Abstract
Purpose: This clinical pilot study was performed to determine the effectiveness of dual-energy cone-beam computed tomography (DE-CBCT) in measuring bone mineral density (BMD).

Materials and Methods: The BMD values obtained using DE-CBCT were compared to those obtained using calibrated multislice computed tomography (MSCT). After BMD calibration with specially designed phantoms, both DE-CBCT and MSCT scanning were performed in 15 adult dental patients. Three-dimensional (3D) Digital Imaging and Communications in Medicine data were imported into a dental software program, and the defined regions of interest (ROIs) on the 3-dimensional surface-rendered images were identified. The automatically-measured BMD values of the ROIs (g/cm3), the differences in the measured BMD values of the matched ROIs obtained by DE-CBCT and MSCT 3D images, and the correlation between the BMD values obtained by the 2 devices were statistically analyzed.

Results: The mean BMD values of the ROIs for the 15 patients as assessed using DE-CBCT and MSCT were 1.09¡¾0.07 g/cm3 and 1.13¡¾0.08 g/cm3, respectively. The mean of the differences between the BMD values of the matched ROIs as assessed using DE-CBCT and calibrated MSCT images was 0.04¡¾0.02 g/cm3. The Pearson correlation coefficient between the BMD values of DE-CBCT and MSCT images was 0.982 (r=0.982, P<0.001).

Conclusion: The newly developed DE-CBCT technique could be used to measure jaw BMD in dentistry and may soon replace MSCT, which is expensive and requires special facilities.
KEYWORD
Bone Mineral Density, Cone-Beam Computed Tomography, Dual Energy, Phantoms