KMID : 1024020200500010023
|
|
Imaging Science in Dentistry 2020 Volume.50 No. 1 p.23 ~ p.30
|
|
Does the metal artifact reduction algorithm activation mode influence the magnitude of artifacts in CBCT images?
|
|
Fontenele Rocharles C.
Nascimento Eduarda H. L. Santaella Gustavo M. Freitas Deborah Q.
|
|
Abstract
|
|
|
Purpose: This study was conducted to assess the effectiveness of a metal artifact reduction (MAR) algorithm activated at different times during cone-beam computed tomography (CBCT) acquisition on the magnitude of artifacts generated by a zirconium implant.
Materials and Methods: Volumes were obtained with and without a zirconium implant in a human mandible, using the OP300 Maxio unit. Three modes were tested: without MAR, with MAR activated after acquisition, and with MAR activated before acquisition. Artifacts were assessed in terms of the standard deviation (SD) of gray values and the contrast-to-noise ratio (CNR) in 6 regions of interest with different distances (10 to 35 mm, from the nearest to the farthest) and angulations (70¡Æ to 135¡Æ) from the implant region.
Results: In the acquisitions without MAR, the regions closer to the implant (10 and 15 mm) had a higher SD and lower CNR than the farther regions. When MAR was activated (before or after), SD values did not differ among the regions (P>0.05). The region closest to the implant presented a significantly lower CNR in the acquisitions without MAR than when MAR was activated after the acquisition; however, activating MAR before the acquisition did not yield significant differences from either of the other conditions.
Conclusion: Both modes of MAR activation were effective in decreasing the magnitude of CBCT artifacts, especially when the effects of the artifacts were more noticeable.
|
|
KEYWORD
|
|
Artifacts, Cone-Beam Computed Tomography, Dental Implant, Zirconium
|
|