KMID : 1024020200500020161
|
|
Imaging Science in Dentistry 2020 Volume.50 No. 2 p.161 ~ p.168
|
|
How image-processing parameters can influence the assessment of dental materials using micro-CT
|
|
Torres Fernanda Ferrari Esteves
Jacobs Reinhilde EzEldeen Mostafa de Faria-Vasconcelos Karla Guerreiro-Tanomaru Juliane Maria dos Santos Bernardo Camargo Tanomaru-Filho Mario
|
|
Abstract
|
|
|
Purpose: The aim of this study was to evaluate the influence of voxel size and different post-processing algorithms on the analysis of dental materials using micro-computed tomography (micro-CT).
Materials and Methods: Root-end cavities were prepared in extracted maxillary premolars, filled with mineral trioxide aggregate (MTA), Biodentine, and Intermediate Restorative Material (IRM), and scanned using micro-CT. The volume and porosity of materials were evaluated and compared using voxel sizes of 5, 10, and 20 ¥ìm, as well as different software tools (post-processing algorithms). The CTAn or MeVisLab/Materialise 3-matic software package was used to perform volume and morphological analyses, and the CTAn or MeVisLab/Amira software was used to evaluate porosity. Data were analyzed using 1-way ANOVA and the Tukey test (P<0.05).
Results: Using MeVisLab/Materialise 3-matic, a consistent tendency was observed for volume to increase at larger voxel sizes. CTAn showed higher volumes for MTA and IRM at 20 ¥ìm. Using CTAn, porosity values decreased as voxel size increased, with statistically significant differences for all materials. MeVisLab/Amira showed a difference for MTA and IRM at 5 ¥ìm, and for Biodentine at 20 ¥ìm. Significant differences in volume and porosity were observed in all software packages for Biodentine across all voxel sizes.
Conclusion: Some differences in volume and porosity were found according to voxel size, image-processing software, and the radiopacity of the material. Consistent protocols are needed for research evaluating dental materials.
|
|
KEYWORD
|
|
Dental Materials, Image Processing, Computer-Assisted, Imaging, Three-Dimensional, X-ray Microtomography
|
|
FullTexts / Linksout information
|
|
|
|