Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1024420120160040333
Food Engineering Progress
2012 Volume.16 No. 4 p.333 ~ p.339
Improvement of Lateral Flow Assay for the rapid detection of Salmonella
Park Saet-Byul

Kim Gi-Young
Moon Ji-Hye
Abstract
In this study, a rapid immuno-chromatographic assay was developed to detect Salmonella Typhimurium. This assay was based on the lateral flow immunoassay. Two different anti-Salmonella polyclonal antibody (Pab) are used in this strip assay for the double antibody sandwich immunoassay method. The strip sensor was composed of sample pad, conjugation pad, membrane, and absorbent pad. For the optimal conjugation between nano-gold and anti-Salmonella Pab, the pH value of colloidal gold solution was determined to 8.4. To prevent non-specific binding between nano gold and anti-Salmonella Pab, sample pad and membrane was treated with PBS buffer, included 2% BSA and 10% sucrose. The control and detection zones were visually detected according to antigen-antibody binding. The detection limit of the developed strip assay was 105 CFU/mL within 10 min, and improved until 104 CFU/mL after 12 hr.
KEYWORD
lateral flow immunoassay, strip sensor, salmonella
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)