Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1024520040130100903
Journal of the Environmental Sciences
2004 Volume.13 No. 10 p.903 ~ p.909
The Optimal Analytical Method for the Determination of PCE and TCE by GC/FID with SPME technieque
An Sang-Woo

Lee Si-Jin
Chang Soon-Woong
Abstract
A new method based on solid phase microextraction(SPME), coupled with GC/FID, has been developed for the determination of PCE and TCE in water samples. The experimental parameters affecting the SPME process (i.e, kinds of fibers, extraction time, desorption time, extraction temperature, volume ratio of sample to headspace, salt addition, and magnetic stirring) were optimized. The coefficients of determination () for PCE and TCE were 0.9951 and 0.9831, respectively when analytes concentration ranges from 10 to 300g/L. The relative standard deviations were 3.4 and for concentration of 10g/L(n=5), respectively. The detection limits of PCE and TCE were 0.5 and l.3g/L, respectively.
KEYWORD
SPME, Headspace analysis, GC-FID, PCE, TCE
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)