Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1028220130250030166
Hip & Pelvis
2013 Volume.25 No. 3 p.166 ~ p.172
Cementless Total Hip Arthroplasty Using the ABG I Hydroxyapatite-coated Prosthesis: Minimum 10 Year Follow-up
Kim Yi-Suk

Park Dong-Hyuck
Hwang Kyu-Tae
Kim Young-Ho
Choi Il-Yong
Abstract
Purpose: The purpose of this study was to evaluate the outcomes of total hip arthroplasty (THA) with an ABG I hip prosthesis after a minimum ten-year follow up.

Materials and Methods: From January 1996 to March 2001, 121 hips in 104 patients, whi were followed up for a minimum of ten-years, were enrolled in this study. The clinical and radiographic outcomes were evaluated and Kaplan-Meier survival analysis was performed.

Results: The mean Harris hip score at the last follow-up was 84 points. A radiolucent line around the cup, osteolysis, and cup loosening were observed in 5 hips(4.1%), 53 hips(43.8%), and 5 hips(4.1%), respectively. In the femoral side, osteolysis and stem loosening were observed in 11 hips(9.1%) and 2 hips(1.7%), respectively. The mean linear wear rate of the polyethylene liner was 0.23 mm/yr. Forty four revisions(36.3%) were performed. The outcomes were similar regardless of the cause of primary THA. The survival rate with the end point of revision due to cup loosening was 56.3%, and revision due to stem loosening was 98.1% after a 16 year follow-up.

Conclusion: High polyethylene wear and the disappointing survival rate of the ABGI cup were observed after a minimum ten year follow-up. Therefore, close observation of patients who have received an ABG I prosthesis is necessary.
KEYWORD
Hip, ABG I, Osteolysis, Polyethylene wear
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed