KMID : 1028220170290040223
Hip & Pelvis
2017 Volume.29 No. 4 p.223 ~ p.227
Results of Total Hip Arthroplasty with 36-mm Metallic Femoral Heads on 1st Generation Highly Cross Linked Polyethylene as a Bearing Surface in Less than Forty Year-old Patients: Minimum Ten-year Results
Choi Won-Kee

Kim Jae-Jung
Cho Myung-Rae
Abstract
Purpose: We evaluated long-term clinical and radiographic results in patients underwent primary total hip arthroplasty (THA) using 36-mm metallic femoral head and 1st generation highly cross-linked polyethylene among patients aged 40 years and less, with minimum 10 years follow-up.

Materials and Methods: We retrospectively reviewed 20 hips underwent primary THA by one hip surgeon using 36-mm femoral head and 1st generation highly cross-linked polyethylene from 2004 to 2006. Mean follow up was 131.55 months. Clinical follow-ups included functional measurements. And we evaluated post operative complications, prosthesis loosening, failure, and osteolysis. For radiologic evaluations, together with position of acetabular cup at 6 weeks later of postoperation, we separately calculated the penentrations of femoral heads into polyethylene liners during postoperation and one year later check-ups, and during one year later check-ups and final check-ups. Polywear pro3D 5.10 software was used to measure penetrations of femoral head.

Results: Mean acetabular cup inclination and anteversion were 49.02¡Æ and 10.19¡Æ, respectively. Mean thickness of the polyethylene liner at 45¡Æ was 6.44 mm. There were no major complications such as implant loosening or osteolysis, and one case of dislocation occurred. Average modified Harris hip score at final follow-up was 91, and Merle d'Aubigne and Postel scores 15 were or over. Average femoral head penetration of bedding wear was 0.170¡¾0.039 mm/year. Steady-state wear rates was 0.059¡¾0.006 mm/year.

Conclusion: Results of THA with 36-mm metallic femoral heads on 1st generation highly cross linked as a bearing surface in less than 40 year-old patients were satisfactory.
KEYWORD
36-mm femoral head, Highly cross-linked polyethylene, Total hip arthroplasty