Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1028220220340020115
Hip & Pelvis
2022 Volume.34 No. 2 p.115 ~ p.121
The Wear Rate and Survivorship in Total Hip Arthroplasty Using a Third-generation Ceramic Head on a Conventional Polyethylene Liner: A Minimum of 15-year Follow-up
Shim Bum-Jin

Park Sung-Jin
Park Chan-Ho
Abstract
Purpose: The purpose of this study was to evaluate the wear and survival rates of third-generation ceramic heads on a conventional ultra-high molecular weight polyethylene liner.

Materials and Methods: A total of 160 hips (147 patients with a mean age of 55.9 years) who underwent total hip arthroplasty using the third-generation ceramic head on a conventional polyethylene liner from March 1998 to August 2003 were reviewed retrospectively. Evaluation of the wear rate for 56 hips (49 patients) followed-up for at least 15 years was performed using the PolyWare program version 8 (Draftware Developers, USA). The Kaplan?Meier survivorship was also evaluated.

Results: Linear wear and volumetric wear rates were 0.11¡¾0.47 mm/year and 32.75¡¾24.50 mm3/year, respectively. Nine revisions were performed during the follow-up period because of cup or stem loosening. The Kaplan?Meier survival rate, using cup revision or total revision total hip arthroplasty (THA) as the endpoint of analysis, was 93.7% at 15 years and 73.6% at 20 years.

Conclusion: Because all revisions were performed between 15 and 20 years in our study, surgeons should pay greater attention to patients who underwent THA with ceramic-on-polyethylene bearing from 15 years postoperatively. Contemporary alumina ceramic on highly cross-linked polyethylene could certainly be a good alternative bearing couple providing better longevity.
KEYWORD
Total hip arthroplasty, Polyethylene, Ceramic, Survivorship