Purpose: Recently, with increased number of patients who require interventional diagnosis or treatment with expanding field of interventional procedures, radiation exposure is of great interest to those who operate interventional procedure. The purpose of our study was to evaluate the radiation dose in table distance and operator distance using Whole body shields for reduction of radiation exposure.
Material and Methods: Philips Allura Xper unit was used with the following protocols: 0.5 §® Pb (60 x 75§¯), collimator (40 x 40§¯ field), SID (90§¯), water phantom (plasty glass, 30 x 40 x 25§¯). Dose measurement was done by using Radical corporation (Radiation monitor controller, model 2026 made in USA) and Electrometer/Ion chamber. Radiation doses were measured 3 times each in 50cm, 100cm and 150cm distance from table and operator, with the described phantom according to the usage of Whole body shields. Measure change on fluoroscopy : R/h ¡æ mR/min ¡æ ¥ìGy/min
Measure change in record dose : mR/min ¡æ ¥ìGy/fps, 1mR = 0.0087mGy
Results: Table 1. Comparision of radiation dose with the usage of shield and distance on fluoroscopy (10 sec)()
Table 2. Comparison of radiation dose with the usage of shield and distance on DSA (2fps)
Conclusion: In comparison of radiation dose with the usage of whole body shield, it demonstrated significant decrease in 50cm distance from table on fluoroscopy (0.03¥ìGy/min ¡æ 0.00¥ìGy/min) and 83% decrease in 50cm distance from operator on DSA (0.02¥ìGy/min ¡æ 0.00¥ìGy/min). However, avoidance of unnecessary fluoroscopy and DSA is needed to reduce radiation exposure to operators, patients and persons related to operation because there is a limitation in perfect radiation shield.
|