KMID : 1034420170420030200
Restorative Dentistry & Endodontics
2017 Volume.42 No. 3 p.200 ~ p.205
In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen
Tampelini Fernanda Garcia

Coelho Marcelo Santos
De Azevedo Rios Marcos
Fontana Carlos Eduardo
Rocha Daniel Guimaraes Pedro
Pinheiro Sergio Luiz
Da Silveira Bueno Carlos Eduardo
Abstract
Objectives: The aim of this in vivo study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF).

Materials and Methods: Thirty-two premolars with single canals that required extraction were included. Following anesthesia, access, and initial canal preparation with size 10 and 15 K-flex files and SX and S1 rotary ProTaper files, the canals were irrigated with 2.5% sodium hypochlorite. The length of the root canal was verified 3 times for each tooth using the 2 apex locators and once using the radiographic technique. Teeth were extracted and the actual WL was determined using size 15 K-files under a ¡¿ 25 magnification. The Biostat 4.0 program (AnalystSoft Inc.) was used for comparing the direct measurements with those obtained using radiographic technique and the apex locators. Pearson's correlation analysis and analysis of variance (ANOVA) were used for statistical analyses.

Results: The measurements obtained using the visual method exhibited the strongest correlation with Root ZX II (r = 0.94), followed by Propex II (r = 0.90) and Ingle's technique (r = 0.81; p < 0.001). Descriptive statistics using ANOVA (Tukey's post hoc test) revealed significant differences between the radiographic measurements and both EALs measurements (p < 0.05).

Conclusions: Both EALs presented similar accuracy that was higher than that of the radiographic measurements obtained with Ingle's technique. Our results suggest that the use of these EALs for MF location is more accurate than the use of radiographic measurements.
KEYWORD
Dental pulp cavity, Electronic apex locators, Endodontic, Radiographs, Therapeutics, Working length