Korean Society for Dental Materials 2017 Volume.44 No. 2 p.151 ~ p.161
|
|
Influence of resin-nanoceramic CAD/CAM block shade and thickness on the microhardness of dual-cured resin cement
|
|
Choi Ga-Young
Park Jeong-Kil Jin Myoung-Uk Kwon Yong-Hoon Son Sung-Ae
|
|
Abstract
|
|
|
The purpose of this study was to examine the effect of shade and thickness of resin-nanoceramic CAD-CAM block (RNB) on the microhardness of dual-cured resin cement, as well as to measure the number of photons transmitted through RNBs of different thicknesses and colors. One dual-cured resin cement was used to prepare resin cement specimens. Resin cement specimens were light-cured for 40 seconds through 3 shades (A1, A2, A3 in HT (high translucency) and LT (low translucency) respectively) and four thicknesses (1, 2, 3, 4 mm) of RNB specimens. Vickers microhardness measurements of resin cement specimens were performed using a Vickers hardness tester. The light transmission of RNB specimens was measured using a spectrometer (SpectroPro-500, Acton Research, Acton, MA, U.S.A.), and the translucency parameter was calculated using the CIEL*a*b* system. Data were statistically analyzed by ANOVA and Tukey?s test. There was a significant decrease of microhardness of resin cement specimen with an overlay of 4 mm of RNB thickness and A3 shade in comparison to A1 and 1 mm, respectively (p<0.05). The translucency parameter values and light transmission of RNBs tested differed significantly, according to the thicknesses of the specimen (p<0.05). Light transmission is decreased with increase in the thicknesses of RNBs. Shade A1 transmitted more light than darker blocks. A decrease in microhardness of resin cement specimens was observed with increasing thickness and shade (A1 to A3) of RNBs.
|
|
KEYWORD
|
|
Resin-nanoceramic CAD/CAM blocks, Dual-cured resin cement, Microhardness
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|