Influence of Band and Loop Type Space Maintainer on Intraoral Scanning Accuracy of an Adjacent Tooth
|
|
Ye Ju-Ri
Chae Yong-Kwon Lee Ko-Eun Lee Hyo-Seol Choi Sung-Chul Nam Ok-Hyung
|
|
Abstract
|
|
|
Purpose: The purpose of this study was to evaluate whether the presence of a space maintainer affects the accuracy of an intraoral scanner.
Materials and Methods: The maxillary primary first molar typodont tooth was removed from the primary dentition typodont model and a band and loop type space maintainer was delivered. After the model was connected to a dental phantom, intraoral scan was performed using TRIOS 4 (3Shape A/S, Copenhagen, Denmark). The scan was repeated with the same technique without the space maintainer. Each scan was performed 10 times. All scan files into a GOM inspect 2018 software and evaluated the accuracy. The accuracy was evaluated on trueness and precision, and calculated using the root mean square value.
Result: When there was a space maintainer in the oral cavity, the trueness value was 0.10¡¾0.02 mm and the precision value was 0.15¡¾0.03 mm. In the absence of the space maintainer, the trueness value was 0.12¡¾0.03 mm and the precision value was 0.16¡¾0.04 mm. There were no significant differences depending on the presence of a space maintainer (P>0.05).
Conclusion: Within the limits of this study, the accuracy of the intraoral scanner was not influenced by the presence of space maintainer.
|
|
KEYWORD
|
|
Children, Dimensional measurement accuracy, Space maintenance
|
|