Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1036620220100021587
Journal of Korean Clinical Health Science
2022 Volume.10 No. 2 p.1587 ~ p.1593
Analysis of refractive error according to the position of vision measurement using an automatic refractor
Kim Bong-Hwan

Han Sun-Hee
An Se-Hyeon
Sin Chang-Seop
Jang Won-Jeong
Hwang In-Seon
Kim Hyung-Soo
Abstract
Purpose. This study was to investigate the degree of refractive error that occurs depending on the measurement location of the subject when performing a refraction test using the automatic refractor.

Methods. When performing the auto-refraction test, measurements were taken while increasing the distance between the forehead and the forehead rest, and the measurements were made by tilting the head clockwise and counterclockwise.
Results. During the auto-refraction test, significant refractive error occurred when the forehead was not attached to the forehead support or the subject's head was turned clockwise or counterclockwise.

Conclusions. When performing a refraction test using an automatic refractor, the examiner will have to pay attention to whether the subject's forehead is in close contact with the forehead rest, and whether the head is tilted.
KEYWORD
Automatic refractor, Measurement position, Refractive error, Vision
FullTexts / Linksout information
Listed journal information