KMID : 1036620220100021587
|
|
Journal of Korean Clinical Health Science 2022 Volume.10 No. 2 p.1587 ~ p.1593
|
|
Analysis of refractive error according to the position of vision measurement using an automatic refractor
|
|
Kim Bong-Hwan
Han Sun-Hee An Se-Hyeon Sin Chang-Seop Jang Won-Jeong Hwang In-Seon Kim Hyung-Soo
|
|
Abstract
|
|
|
Purpose. This study was to investigate the degree of refractive error that occurs depending on the measurement location of the subject when performing a refraction test using the automatic refractor.
Methods. When performing the auto-refraction test, measurements were taken while increasing the distance between the forehead and the forehead rest, and the measurements were made by tilting the head clockwise and counterclockwise.
Results. During the auto-refraction test, significant refractive error occurred when the forehead was not attached to the forehead support or the subject's head was turned clockwise or counterclockwise.
Conclusions. When performing a refraction test using an automatic refractor, the examiner will have to pay attention to whether the subject's forehead is in close contact with the forehead rest, and whether the head is tilted.
|
|
KEYWORD
|
|
Automatic refractor, Measurement position, Refractive error, Vision
|
|
FullTexts / Linksout information
|
|
|
|