Reconsidering the ¡°MR Unsafe¡± breast tissue expander with magnetic infusion port: A case report and literature review
Dibbs Rami

Culo Bozena
Tandon Ravi
Shellock Frank G.
Lau Frank H.
Abstract
Breast tissue expanders (TEs) with magnetic infusion ports are labeled ¡°MR Unsafe.¡± Therefore, patients with these implants are typically prevented from undergoing magnetic resonance imaging (MRI). We report a patient with a total submuscular breast TE who inadvertently underwent an MRI exam. She subsequently developed expander exposure, requiring explantation and autologous reconstruction. The safety profile of TEs with magnetic ports and the use of MRI in patients with these implants is surprisingly controversial. Therefore, we present our case report, a systematic literature review, and propose procedural guidelines to help ensure the safety of patients with TEs with magnetic ports that need to undergo MRI exams.
KEYWORD
Breast tissue expander, Breast reconstruction, Magnetic resonance imaging, Safety
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø