The response properties of continuous automated system using an enzyme reactor for determination of urea were investigated. The enzyme reactor was constructed to packed-bed form which filled with nylon-6 beads (42¡48 mesh), which immobilized urease with glutaraldehyde, in teflon tube (2 mm I.D., 20 cm length). The system was composed of the enzyme reactor, gas dialyzer, and tublar PVC-nonactin membrane ammonium ion-selective electrode as an indicator electrode in serial order. The response characteristics of this system were as follows. That is, the concentration range of linear response, slope of linear response, detection limit, and conversion percentage were 5.5 ¡¿ 10-6¡2.4 ¡¿ l0-3M, 57.8 mV/decade, 1.5 ¡¿ 10-6M, and 80.8%, respectively. The optimum buffer and life time of urease reactor were 0.01M Tris-HCl buffer solution (pH 7.0¡7.8) and 0.01M phosphate buffer solution (pH 6.9¡7.5) and about 150 days, respectively. And the urease reactor had no interferences of the other physiological materials.
|