Development of Portable Gas Chromatography / Photoionization Detector System
|
|
Kim Man-Goo
Shim Jee-Hee Lee Dong-Soo Lee Yong-Keun
|
|
Abstract
|
|
|
The portable gas chromatography system was developed which was consisted of ambient vapor sampler(AVS), short capillary column(3 m long, 0.32 mm i.d. GC(SCCGC), photoionization detector (PID) and vacuum pump which was operated at subambient pressure. The seletion of capillary column was based on the theoretical calculation from Golay equation. The pressure ratio of column inlet and outlet appropriated between 1.03 and 1.2 in the system. The available column flow were 0.87¡4.63 ml/min at the pressure ratios. The AVS consisted of three concentric tubes and enables rapid, repetitive introduction of vapor samples directly into capillary column and showed good reproducibility. The subambient column outlet pressure with PID resulted in a significant increase in the optimum column flow, permitting rapid analysis. The baseline separation of m-xylene and o-xylene was able to within 40 second with the system. Parameters affecting the column resolving power were sampling duration, column length and diameter, and the pressure ratio. Effects of these parameters were investigated using bezene derivative compounds.
|
|