The analysis of trace cadmium in blood by using 10 folds diluted samples with 1% Triton X-100 and Fork platform tube at the ashing temperature of 500¡É has been performed by GFAAS. In this study, good analytical results were obtained from peak area and peak height methods. The detection limits of SH and D2 arc background correction are 0.02 ng/mL and 0.01 ng/mL, respectively. Relative standard deviation is within 5% at the level of 1.0 ng/mL.
|
|