Journal of the Korean Chemical Society 1997 Volume.41 No. 6 p.292 ~ p.298
|
|
Development of an Axially Viewed Inductively Coupled Plasma for Atomic Emission Spectrometry and Comparison between the Detection Limits of Lead
|
|
Cho Sung-il
Han Myung-Sub Lee Sang-Hwa Lee Joung-Hae Woo Jin-Chun
|
|
Abstract
|
|
|
An ICP(Inductively Coupled Plasma) emission spectrometer was developed with an axially viewed ICP source incorporated by a 5-turned induction coil and a torch, outer quartz tube of which was 50 mm longer than that used in conventional ICP/AES(Inductively Coupled Plasma Atomic Emission Spectrometry). The Optimization of the system has been performed in terms of the determination of signal-to-noise ratio and background intensity at various rf powers, sample flow rates, argon gas flow rates and cut-off gas flow rates. The spectro-analytical characteristics of the spectrum obtained between 200 and 500 nm was revealed to be similar compared with a vertically viewed ICP source. The detection limit of Pb(¥±) at 220.35 nm was 11 ppb which was 5 times lower than that obtained with a vertically viewed ICP source.
|
|