The new GD/ICP-AES quick change over system has been developed and characterized. Within less than 15 minutes, ICP-AES could be switched to GD-AES and vise a versa. As a result, both solid and liquid samples could be analyzed in a very short period of time by the ICP/GD-AES quick change over system developed in our laboratory. The influences of the experimental variables, such as flow rate of coolant gas, flow rate of auxiliary gas, flow rate of sample carrier gas, sampling depth, orifice size of sampling cone, and rf (radio frequency) power on emission intensity have been presented. The detection limits of Cd(I) 228.8 nm, Mn (II) 257.61 nm, and Fe(II) 259.95 nm were found to be 3.86, 1.49, and 5.79 ppb, respectively. And linealities of the calibration curves were measured to be unity.
|