Journal of the Korean Chemical Society
1999 Volume.43 No. 4 p.407 ~ p.411
Determination of Cr(¥µ) by Glassy Carbon and Platinum Electrodes Modified With Polypyrrole Film
Yoo Kwang-Sik

Woo Sang-Beom
Jyoung Jy-Young
Abstract
Studies have been carried out on the fabrication of PPy/GC and PPy/Pt electrode modified with polypyrrole film and determination of Cr(VI) by using 3-electrode system with modified electrodes. Modified electrodes were able to easily fabricated by cyclic voltammetry scanned from +1.0V to -1.0V(vs. Ag/AgCl) at 50 mV/sec. Film thickness could be controlled at same condition by the number of cycling up to 26 times. Reduction behaviour of Cr(VI) at PPy/GC electrode could be seen at wide potential ranges from +0.6V to -0.5V(vs. Ag/AgCl), and maximum reduction peak potential of the ion was observed at -0.25V(vs.Ag/AgCl). Calibration graph at its potential was linear from 0.1 ppm to 80.O ppm. Slope factor and relative coefficient were 1.75 mA/ppm and 0.998, respectively. Reduction behaviour of Cr(VI) at PPy/Pt electrode was similar to PPy/GC electrode, Calibration graph was linear from l.0 ppm to 60.0 ppm. Slope factor and relative coefficient were 0.5mA/ppm and 0.923, respectively. But PPy/GC modified electrode had about 3 times higher sensitivity than PPy/Pt modified electrode. Reduction behaviour of Cu(II), As(IlI), Pb(II), and Cd(II) couldn't be seen at PPy/GC electrode,Its metals had not lnterfered with Cr (VI) determination.