Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1059520060500030208
Journal of the Korean Chemical Society
2006 Volume.50 No. 3 p.208 ~ p.215
Real-time Pesticide Assay on Live Tissue Using Electrochemical Graphite Pencil Electrode
À̼ö¿µ/Ly SY
Abstract
A simply prepared graphite, pencil-type working electrode was utilized to monitor fenitrothion concentrations, using the cyclic voltammetry (CV) and square-wave (SW) stripping voltammetry methods. The optimum conditions for analysis were sought. A very low detection limit was obtained compared to that obtained when other common voltammetry methods are used. The optimal parameters of the pencil-type electrode were found to be as follows: a pH of 3.7, a frequency of 500 Hz, an SW amplitude of 0.1 V, an increment potential of 0.005 V, an initial potential of -0.9 V, and a deposition time of 500 sec. The analytical detection limit was determined to be 6.0 ngL-1 (2.164¡¿10-11 molL-1) fenitrothion at SW anodic and CV, and the relative standard deviation at the fenitrothion concentration of SW anodic 10 ugL-1 was 0.30% (n = 15) under the optimum conditions. Analysis was directly conducted through in-vivo realtime assay.
KEYWORD