Journal of the Korean Chemical Society
2019 Volume.63 No. 4 p.246 ~ p.252
A Study on the New Development for Determination of Dead Time in GC-OTC/FID
Oh Doe-Suk

Kim Sung-Hwa
Ko Eun-Ah
Jeon Hyung-Woo
Abstract
In the system of GC-OTC/FID (Gas chromatography-Open Tubular Column/Flame Ionization Detector), DMSO (Dimethyl sulfide) solvent was used to separate the polar solvents (Alcohols). In this system DMSO was eluted later than the separated polar solvents. At this system to calculate chromatographic factors [adjusted retention time (tR'=tR-tO), capacity factor{ k'=(tR-tO)/tO} and separation factor {¥á=(tR2-tO)/(tR1-tO)}], dead time(tO) is necessary, but the method to calculate it has not been reported yet. Therefore, we have tried to develop tO. To calculate tO, we conversed DMSO retention time (DMSO tR) to logarithm (f(x)=log tR(DMSO)¡ætO, tO=log 9.551=0.980). To confirm the optimization of the developed method, we compared with CH4 tR and ln tR(DMSO). Both of the values calculated by CH4 tR and ln tR(DMSO) were not suitable in the calculation k' and ¥á. The developed method in this study{log tR(DMSO)} has satisfied both of the values k' criteria (1
KEYWORD
Chromatographic factors, Dead time (tO), DMSO, Logarithm