Performance Evaluation of a Portable GC for Real-time Monitoring of Volatile Organic Compounds
|
|
You Dong-Wook
Seon Yeong-Sik Oh Jun-Sik Yi Bong-Yoon Kim Hyun-Sik Jung Kwang-Woo
|
|
Abstract
|
|
|
Performance of a portable GC that can be utilized for the real time determination of volatile organic compounds in air was evaluated. It employs purified/compressed ambient air as the carrier gas eliminating the need for high pressure gas tanks. The compact system with dimensions of 35 ¡¿ 26 ¡¿ 15 cm3 and weight of 5 kg is powered by either a 24 V DC external adapter or battery pack. Chromatograms of the mixture sample including benzene, toluene, ethylbenzene, and oxylene at concentrations of 1 ppmv and 20 ppmv represent a good reproducibility: 3.79% and 0.48% relative standard deviations (RSDs) for peak area variations; 0.40% and 0.08% RSDs for retention times. The method detection limit was 0.09 ppmv. A 30 m long, 0.28 mm I.D. column operated at an optimal condition yielded a peak capacity of 61 with good resolution for a 10 min isothermal analysis. The relative standard deviations (RSD) of the peak area variations and retention times during consecutive measurements over 27 h were less than 2.4%RSD and 0.5%RSD, respectively. Thus, this instrument makes it suitable for continuous and field analysis of low-concentration VOC mixtures in the indoor/outdoor environment as well as the spillage accident of hazardous chemicals.
|
|
KEYWORD
|
|
Portable GC, Reproducibility, Detection limit, Resolution, Volatile organic compounds
|
|