KMID : 1100720190390050478
|
|
Annals of Laboratory Medicine 2019 Volume.39 No. 5 p.478 ~ p.487
|
|
Comparison of cobas EGFR Mutation Test v2 and PANAMutyper-R-EGFR for Detection and Semi-Quantification of Epidermal Growth Factor Receptor Mutations in Plasma and Pleural Effusion Supernatant
|
|
Han A-Lum
Kim Hak-Ryul Choi Keum-Ha Hwang Ki-Eun Zhu Mengyu Huang Yuya Wu Moxin Lee Young-Jin Park Min-Cheol Cho Ji-Hyun Park Do-Sim
|
|
Abstract
|
|
|
Background: Plasma epidermal growth factor receptor (EGFR) mutation tests are less invasive than tissue EGFR mutation tests. We determined which of two kits is more efficient: cobas EGFR Mutation test v2 (cobasv2; Roche Molecular Systems, Pleasanton, CA, USA) or PANAMutyper-R-EGFR (Mutyper; Panagene, Daejeon, Korea). We also evaluated whether pleural effusion supernatant (PE-SUP) samples are assayable, similar to plasma samples, using these two kits.
Methods: We analyzed 156 plasma and PE-SUP samples (31 paired samples) from 116 individuals. We compared the kits in terms of accuracy, assessed genotype concordance (weighted ¥ê with 95% confidence intervals), and calculated Spearman's rho between semi-quantitatively measured EGFR-mutant levels (SQIs) measured by each kit. We also compared sensitivity using 47 EGFR-mutant harboring samples divided into more-dilute and less-dilute samples (dilution ratio: ¡Ã or <1:1,000).
Results: cobasv2 tended to have higher accuracy than Mutyper (73% vs 69%, P=0.53), and PE-SUP samples had significantly higher accuracy than plasma samples (97% vs 55?71%) for both kits. Genotype concordance was 98% (¥ê=0.92, 0.88?0.96). SQIs showed strong positive correlations (P<0.0001). In less-dilute samples, accuracy and sensitivity did not differ significantly between kits. In more-dilute samples, cobasv2 tended to have higher sensitivity than Mutyper (43% vs 20%, P=0.07).
Conclusions: The kits have similar performance in terms of EGFR mutation detection and semi-quantification in plasma and PE-SUP samples. cobasv2 tends to outperform Mutyper in detecting less-abundant EGFR-mutants. PE-SUP samples are assayable using either kit.
|
|
KEYWORD
|
|
Epidermal growth factor receptor, Mutation, Plasma, Pleural effusion supernatant, cobas EGFR Mutation test v2, PANAMutyper-R-EGFR
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|