KMID : 1100720220420030331
|
|
Annals of Laboratory Medicine 2022 Volume.42 No. 3 p.331 ~ p.341
|
|
Effect of Outlier Elimination on the 99th Percentile Upper Reference Limits of High-Sensitivity Cardiac Troponin I Assays Based on a Strictly Selected Healthy Reference Population
|
|
Ahn Sun-Young
Kim Hyun-Ki Lee Woo-Chang Chun Sa-Il Min Won-Ki
|
|
Abstract
|
|
|
Background: We established high-sensitivity cardiac troponin I (hsTnI) 99th percentile upper reference limits (URLs) for the Centaur XPT High-Sensitivity Troponin I assay (Centaur hsTnI; Siemens, Erlangen, Germany) and Atellica IM High-Sensitivity Troponin I assay (Atellica hsTnI; Siemens) and assessed the effect of outlier elimination.
Methods: The reference population comprised 380 men and 387 women, satisfying the strict systematic reference population criteria. After reference population verification by the N-terminal pro-B-type natriuretic peptide (NT-proBNP) assay, 99th percentile URLs for Centaur hsTnI and Atellica hsTnI were calculated before and after outlier elimination.
Results: The 99th percentile URL for Centaur hsTnI was 60.4 (men, 74.7; women, 57.5) ng/L and that for Atellica hsTnI was 59.6 (men, 75.2; women, 55.1) ng/L. After the elimination of 61 (8.0%) outlier samples in Centaur hsTnI and 58 (7.6%) in Atellica hsTnI, the 99th percentile URLs were 13.5 ng/L (men, 15.3 ng/L; women, 11.9 ng/L) and 13.4 ng/L (men, 15.5 ng/L; women, 12.9 ng/L), respectively, significantly lower than those before outlier elimination. The CVs at the 99th percentile URLs were 5.2% and 3.5%, respectively. The measurable fractions among the reference population were 91.5% and 93.4%, respectively. Performance evaluation of Atellica B-type natriuretic peptide (BNP), Atellica NT-proBNP, Centaur hsTnI, and Atellica hsTnI showed outstanding results.
Conclusions: The Korean hsTnI 99th percentile URLs calculated in this study were significantly lower after outlier elimination than before. Centaur hsTnI and Atellica hsTnI meet the ¡°Guideline acceptable¡± and ¡°Level 3 (second generation, high sensitivity)¡± requirements, satisfying international standards.
|
|
KEYWORD
|
|
Ninety-ninth percentile, Upper reference limit, Korean, Outlier, ADVIA Centaur XPT High-Sensitivity Troponin I assay, Atellica IM High-Sensitivity Troponin I assay
|
|