Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1100720220420040398
Annals of Laboratory Medicine
2022 Volume.42 No. 4 p.398 ~ p.405
Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time
Nam Min-Jeong

Yoon Su-Mi
Hur Mi-Na
Lee Gun-Hyuk
Kim Han-Ah
Park Mi-Kyoung
Kim Hyeong-Nyeon
Abstract
Background: Digital morphology (DM) analyzers are increasingly being used for white blood cell (WBC) differentials. We assessed the laboratory efficiency of the Sysmex DI-60 system (DI-60; Sysmex, Kobe, Japan) in comparison with manual counting in leukopenic samples.

Methods: In total, 40 peripheral blood smear samples were divided into normal, mild leukopenia, moderate leukopenia, and severe leukopenia groups based on WBC count. In each group, the risk and turnaround time (TAT) were compared between DI-60 and manual counting. Risk was determined by failure mode and effect analysis using the risk priority number (RPN) score, and TAT was recorded for the analytical phase.

Results: Overall, DI-60 showed a five-fold lower risk (70 vs. 350 RPN) and longer TAT than manual counting. In severe leukopenic samples, DI-60 showed a shorter TAT/slide and a remarkably lower cell count/slide than manual counting. In all samples, the TAT/cell for DI-60 was substantially longer than that for manual counting (DI-60 vs. manual: total, 1.8 vs. 1.0 sec; normal, 1.5 vs. 0.7 sec; mild leukopenia, 1.9 vs. 0.9 sec; moderate leukopenia, 1.8 vs. 1.0 sec; severe leukopenia, 28.8 vs. 19.0 sec).

Conclusions: This is the first comparative assessment of risk and TAT between DI-60 and manual counting in leukopenic samples. DI-60 decreases the laboratory risk and improves patient safety, but requires more time to count fewer cells, especially in severe leukopenic samples. DM analyzers should be applied selectively depending on the WBC count to optimize laboratory efficiency.
KEYWORD
Leukopenia, DI-60, WBC differential, Risk, Turnaround time, Laboratory efficiency
FullTexts / Linksout information
Listed journal information
SCI(E) MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed