Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1101120200170020148
Journal of Sleep Medicine
2020 Volume.17 No. 2 p.148 ~ p.153
Is Split-Night Polysomnography Valid for Women with Obstructive Sleep Apnea?
Kang Mi-Ri

Ji Ki-Hwan
Abstract
Objectives: Split-night polysomnography (PSG) can be used for the diagnosis and treatment of obstructive sleep apnea (OSA). Recently, the American Academy of Sleep Medicine has broadened the indication of split-night PSG from severe to moderate OSA in a time window of 2 hours. However, majority pieces of evidence have been derived from the male predominant study population. The objective of this study was to investigate the diagnostic accuracy of split-night PSG in Korean women.

Methods: This study included 122 adult women with OSA. Apnea-hypopnea index (AHI) from the first 120 minutes of clock time (T120-AHI) and full-night AHI (FN-AHI) were compared using the concordance correlation coefficient (CCC) method and Bland-Altman plot. Receiver operating characteristic (ROC) curves and the area under the ROC curves were plotted with various cut-off points of AHI.

Results: The T120-AHI correlated with FN-AHI (CCC=0.77). The area under the ROC curve for T120-AHI with FN-AHI ¡Ã15 was 0.865, with a sensitivity of 78.6% and specificity of 82.7%.

Conclusions: T120-AHI showed a significant correlation with the FN-AHI value in Korean female patients with OSA.
KEYWORD
Split-night polysomnography, Polysomnography, Woman, Obstructive sleep apnea, Daignosis