Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1101320060380030147
Korean Journal of Clinical Laboratory Science
2006 Volume.38 No. 3 p.147 ~ p.151
The Effect of the Storage Duration and Temperature of EDTA Specimen for CBC and WBC Differential Count in SE-9000 Automated Cell Counter
Hong Seung-Bok

Kim Jong-Seok
Shin Kyeong-Seob
Abstract
Although various automated CBC analyzers with different WBC analytical principles were consequently introduced to clinical laboratory, the specific information concerning the suitability or unsuitability of aging samples is scarce. For this reason, we studied the effect of storage duration and temperature on CBC parameter in SE-9000 (SYSMEX Medical Electronics Co., Ltd., Kobe, Japan), automated CBC analyzer. We tested 32 K3-EDTA specimens with SE-9000 during 72 hours. Specimens were kept at room temperature (RT) and refrigerated and were analyzed at 0 hr, 4 hr, 8 hr, 24 hr, 48 hr and 72 hr after the collection of the specimens. The percentage changes from initial value for each parameters were calculated. Among the CBC parameters, hemoglobin, red blood cell count, mean corpuscular hemoglobin and platelets were stable for the study period at both temperatures. The mean corpuscular volume (MCV), hematocrit (Hct) and red cell distribution (RDW) increased and the mean corpuscular hemoglobin concentration (MCHC) decreased over time at room temperature. These parameters were stable when refrigerated. The leukocyte count was stable during 72hr at RT and when refrigerated. At room temperature, the relative percentages of neutrophils tend to increase, whereas those of lymphocyte and monocytes tend to decrease after 48 hours. When refrigerated, those of neutrophils and monocytes tend to increase, whereas those of lymphocytes tend to decreased over time. CBC parameters of refrigerated specimen were reliable for 72 hr for the exception of differential count from 24 hr but many CBC parameters, such as MCV, Hct, MCHC, RDW and differential count of leukocyte of blood stored at room temperature for 24 hr were unreliable.
KEYWORD
CBC parameter, Storage, Room temperature, Automated cell counter
FullTexts / Linksout information
Listed journal information