Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1101320090410010042
Korean Journal of Clinical Laboratory Science
2009 Volume.41 No. 1 p.42 ~ p.46
Evaluation of Clinical Utility and Biologic False Positive (BFP) Rates in Automated Syphilis Test Kits for Syphilis Screening
Kim Sung-Man

Lee Je-Hoon
Abstract
Unlike most bacteria, Treponema pallidum subspecies cannot be readily isolated or sustained in cell culture for numerous generations. In korea, two non treponemal tests are currently considered as standard; the VDRL slide test and RPR card test. These tests are based on an antigen composed of an alcoholic solution containing measured amount of cardiolipin, cholesterol, and sufficient purified lecithin to produce reactivity. The nontreponemal reagin tests measure immunoglobulin M (IgM) and IgG to lipoidal material released from damaged host cells as well as to lipoprotein-like material and possibly by cardiolipin released from the treponemes. The object of the evaluation was to evaluate the performance of the Mediace RPR kit on the automated biochemistry analyzer system as a method for screen method of syphilis as well as to identify BFP possibility. For evaluation of routine screening test, a total 2,380 specimens tested by Mediace RPR from 28th Oct, 2007 to 22th Feb, 2008. For evaluation of BFP possiblility, we measured samples which have potential BFP reaction in Syphilis test such as ANA (anti-nuclear antibody) positive (135 samples), CRP (C-reactive protein) positive (100 samples), RF (Rheumatoid factor) positive (26 samples), and other potential BFP cases (17 samples) including total 278 samples. These samples were tested quantitative test Mediace RPR with Hitachi 7600 P module. For comparison with current manual test, VDRL slide test were performed. Of these 2380 specimens, 2350 were negative, 30 were positive, and one were positive with TPHA. Both methods agreed for 2356 (98.9%) samples. Of the 30 samples showed positive results over 1.0 R.U, 6 samples showed positive results with VDRL test. Of these 6 samples, 1 samples showed positive with TPHA test. The combination of the Automated Biochemistry analyzer and VDRL test for retest can be increase efficiency of syphilis screening test.
KEYWORD
Mediace RPR, VDRL, TPHA, BFP
FullTexts / Linksout information
Listed journal information