KMID : 1102020160460030155
|
|
Applied Microscopy 2016 Volume.46 No. 3 p.155 ~ p.159
|
|
Relationship of the Distribution Thickness of Dielectric Layer on the Nano-Tip Apex and Distribution of Emitted Electrons
|
|
Al-Qudah Ala¡¯a M.
Mousa Marwan S.
|
|
Abstract
|
|
|
This paper analyses the relationship between the distribution of a dielectric layer on the apex of a metal field electron emitter and the distribution of electron emission. Emitters were prepared by coating a tungsten emitter with a layer of epoxylite resin. A high-resolution scanning electron microscope was used to monitor the emitter profile and measure the coating thickness. Field electron microscope studies of the emission current distribution from these composite emitters (Tungsten-Clark Electromedical Instruments Epoxylite resin [Tungsten/CEI-resin emitter]) have been carried out. Two forms of image have been observed: bright single-spot images, thought to be associated with a smooth substrate and a uniform dielectric layer; and multi-spot images, though to be associated with irregularity in the substrate or the dielectric layer.
|
|
KEYWORD
|
|
Field electron emitter, Dielectric layer, Composite emitter, Field electron microscope, Scanning electron microscope
|
|