Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1103720070560040315
Journal of the Korean Society of Radiology
2007 Volume.56 No. 4 p.315 ~ p.320
The Safety and Efficacy of Hemostasis with Clo-Sur P.A.D. after Transcatheter Arterial Chemoembolization.
Jung Seung-Chai

Lee Whal
Park Jae-Hyung
Jae Hwan-Jun
Kim Sang-Youn
Chung Jin-Wook
Abstract
Purpose: We wanted to evaluate the safety and efficacy of a new hemostatic device, Clo-Sur P.A.D., at an arterial access site after performing femoral arterial catheterization to achieve transcatheter arterial chemoembolization (TACE).

Materials and Methods:From August 2002 to March 2005, 113 patients who underwent TACE and agreed on using the Clo-Sur P.A.D. were enrolled in this study. We evaluated the mean time interval from compression to the first movement and also to the first walk. We also evaluated such complications as rebleeding, pseudoaneurysm, vascular occlusion, hematoma, infection and pain. For 92 patients who had previous experiences with manual compression, we evaluated their preference of hemostatic method by asking them.

Results:Successful hemostasis was achieved with the Clo-Sur P.A.D. in 105 subjects (92%). The mean time interval from compression to the first movement was 201 minutes, and that to the first walk was 267 minutes. There was no statistical difference between the complicated and the uncomplicated groups for the mean time, prothrombin time, Child-Pugh class and platelet count (p > 0.05). Rebleeding occurred in 3 patients (2.7%) and mild hematoma around puncture site was noted in 5 patients (5.4%). Eighty-seven patients (95%) preferred Clo-Sur P.A.D. to the manual compression method.

Conclusions: The Clo-Sur P.A.D. is a safe and effective hemostatic device and it provides early ambulation after TACE.