Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1103720100630040351
Journal of the Korean Society of Radiology
2010 Volume.63 No. 4 p.351 ~ p.357
Placement of a Hemodialysis Catheter using the Dilated Right External Jugular Vein as a Primary Route
Park Mi-Hyun

Shin Byung-Seok
Abstract
Purpose: To evaluate the feasibility that a dilated right external jugular vein (EJV) could be a primary venous access site for large bore hemodialysis catheter placement.

Methods:Between January 2008 and April 2009, a total of 173 hemodialysis catheters (14.5 F) were placed. Among them, we evaluated the clinical data of 42 patients who underwent placement through a dilated right EJV. We evaluated technical success, duration of catheterization in days, and the presence of complications.

Results:Technical success was achieved for 41 patients (98%). Catheter placement was unsuccessful in one patient due to narrowing of the EJV. The catheter dwell time ranged between 14 and 305 days (mean; 76 days, total catheter days: 3,111 days). A total of 26 hemodialysis catheters were removed due to complications (n=2) and termination of hemodialysis via the hemodialysis catheter (n=24). There was air embolization (n=1) and catheter kinking (n=3) during procedures and catheter related infections (n=2) during the follow-up period. The incidence of catheter related infection was 0.06 per 100 catheter days. No cases of catheter malfunction or symptomatic venous thrombosis were observed.


Conclusions: We suggest that a dilated right EJV could be considered as a preferred primary route for hemodialysis catheter placement with easy access.
KEYWORD
Catheterization, Catheters, Indwelling, Jugular Veins, Renal Dialysis
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø