Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1103720110650060555
Journal of the Korean Society of Radiology
2011 Volume.65 No. 6 p.555 ~ p.562
Factors Affecting Longevity of Tunneled Central Venous Cathe
Park Ji-Won

Lee Jong-Min
Abstract
Purpose: To evaluate factors which affect the longevity of tunneled central venous catheters (T-CVCs).

Materials and Methods : A retrospective study was conducted on 363 T-CVCs. We evaluated the relevant factors affecting the longevity of the T-CVCs, such as age, gender, indication for catheterization, site of entry vessel, diameter and type of T-CVC, catheter tip position, and underlying diseases.

Results: Of the 363 T-CVCs which had been inserted, 331 (91%) were placed through the right internal jugular vein (RIJV). The catheter tip position was the strongest predictor for the longevity of the T-CVC. The short limb of the catheter tip placed either at the cavoatrial junction (CA junction) or below the lower margin of the right main bronchus had a good prognosis. The vessel through which the T-CVC was placed significantly influenced the longevity of the T-CVC; the RIJV was associated with better results than the left internal jugular vein. Also, a split-type catheter was significantly associated with a better result.

Conclusion: A two distinct and separate type T-CVC placed through the RIJV in which the short limb catheter tip position was at the level of the CA junction, significantly increased the longevity of T-CVCs.
KEYWORD
Tunneled Central Venous Catheters, Longevity, Hemodialysis, Internal Jugular Vein, Catheter Tip
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø