KMID : 1103720170760030173
Journal of the Korean Society of Radiology
2017 Volume.76 No. 3 p.173 ~ p.178
Comparison of Image Quality between Mammography Dedicated Monitor and UHD 4K Monitor, Using Standard Mammographic Phantom: A Preliminary Study
Lee Ji-Young

Cha Soon-Joo
Hong Sung-Hwan
Kim Su-Young
Kim Yong-Hoon
Kim You-Sung
Kim Jeong-A
Abstract
Purpose: Using standard mammographic phantom images, we compared the image quality obtained between a mammography dedicated 5 megapixel monitor (5M) and a UHD 4K (4K) monitor with digital imaging and communications in medicine display, to investigate the possibility of clinical application of 4K monitors.

Materials and Methods: Three different exposures (autoexposure, overexposure and underexposure) images of mammographic phantom were obtained, and six radiologists independently evaluated the images in 5M and 4K without image modulation, by scoring of fibers, groups of specks and masses within the phantom image. The mean score of each object on both monitors was independently analyzed, using t-test and interobserver reliability by intraclass correlation coefficient (ICC) of SPSS.

Results: The overall mean scores of fiber, group of specks, and mass in 5M were 4.25, 3.92, and 3.28 respectively, and scores obtained in 4K monitor were 3.81, 3.58, and 3.14, respectively. No statistical difference was seen in scores of fiber and mass between the two monitors at all exposure conditions, but the score of group of specks in 4K was statistically lower in the overall (p = 0.0492) and in underexposure conditions (p = 0.012). The ICC for interobserver reliability was excellent (0.874).

Conclusion: Our study suggests that since the mammographic phantom images are appropriate with no significant difference in image quality observed between the two monitors, the 4K monitor could be used for clinical studies. Since this is a small preliminary study using phantom images, the result may differ in actual mammographic images, and subsequent investigation with clinical mammographic images is required.
KEYWORD
Mammography, Phantoms, Imaging, Radiologists, Radiology Information Systems