KMID : 1103720170770020089
Journal of the Korean Society of Radiology
2017 Volume.77 No. 2 p.89 ~ p.96
Effective Dose in Abdominal Digital Radiography: Patient Factors
Jang Ji-Sung

Koo Hyun-Jung
Yang Hyung-Jin
Park Jung-Hoon
Cho Young-Chul
Do Kyung-Hyun
Abstract
Purpose: To identify independent patient factors associated with an increased radiation dose, and to evaluate the effect of patient position on the effective dose in abdominal digital radiography.

Materials and Methods: We retrospectively evaluated the effective dose for abdominal digital radiography in 222 patients. The patients were divided into two groups based on the cut-off dose value of 0.311 mSv (the upper third quartile of dose distribution): group A (n = 166) and group B (n = 56). Through logistic regression, independent factors associated with a larger effective dose were identified. The effect of patient position on the effective dose was evaluated using a paired t-test.

Results: High body mass index (BMI) (¡Ã 23 kg/m2), presence of ascites, and spinal metallic instrumentation were significantly associated with a larger effective dose. Multivariate logistic regression analysis revealed that high BMI [odds ratio (OR), 25.201; p < 0.001] and ascites (OR, 25.132; p < 0.001) were significantly associated with a larger effective dose. The effective dose was significantly lesser (22.6%) in the supine position than in the standing position (p < 0.001).

Conclusion: High BMI and ascites were independent factors associated with a larger effective dose in abdominal digital radiography. Significant dose reduction in patients with these factors may be achieved by placing the patient in the supine position during abdominal digital radiography.
KEYWORD
Radiation Dosage, Abdominal Radiography, Digital Radiography