KMID : 1103720190800020306
Journal of the Korean Society of Radiology
2019 Volume.80 No. 2 p.306 ~ p.317
Medical Institutions' Adherence to the PI-RADS v2 Minimum Technical Standards for Prostate MRI in Korea
Kwon Young-Woo

Choi Moon-Hyung
Jung Seung-Eun
Abstract
Purpose: To evaluate the adherence rate to Prostate Imaging-Reporting and Data System version 2 (PI-RADS v2) minimum technical standards of prostate magnetic resonance imaging (MRI) in Korean medical institutions.

Materials and Methods: This study included 103 prostate MRI examinations from 85 outside medical institutions performed from March 2015 to January 2018. The difference in adherence rate to minimal technical standards of PI-RADS v2 was compared using a Fisher's exact test between subgroups divided by the magnetic strength of MRI machine, type of medical institution and presence of genitourinary radiologist.

Results: Diffusion-weighted imaging (DWI) was obtained frequently in examinations performed in a 3-T machine, in university hospitals and in medical institutions where genitourinary radiologist work in than the others (p < 0.001, p < 0.001, p = 0.003). Many minimum technical standards of PI-RADS v2 showed significantly lower adherence rate in a 1.5-T machine, in a non-university hospital and in a medical institution without genitourinary radiologist than the others.

Conclusion: The frequency of obtaining DWI and the adherent rate to some of the PI-RADS v2 minimum technical standards were significantly higher in 3-T machines, university hospitals and medical institutions with a genitourinary radiologist.
KEYWORD
Magnetic Resonance Imaging, Prostate, Standards