Journal of Korean Society of Computed Tomographic Technology
2016 Volume.18 No. 2 p.41 ~ p.47
The Study on the Useful Evaluation to the Reduction of Exposure Dose of Urograpy CT through the BMI
Kim Ji-Nil

Park Seong-Hyeok
Jung Hyeong-Sik
Abstract
Purpose : In this study, we tried to investigate the usefulness of relationship between BMI and patient dose by varying exposure conditions, for the goal of X-ray dose reduction of patients.

Materials and Methods : We grouped 156 patients in the age of 20 to 70 who performed Urography CT from Jan 2015 to Dec 2015 into three groups A, B, and C to measure and analyze both DLP and BMI values. Group A consists of 52 patients who took 100kV manual exposure control method from Jan 2015 to Jun 2015, Group B consists of 52 patients who took 100kV automatic exposure control from Jun 2015 to Oct 2015, and Group C consists 52 patients who took 100kV/120kV automatic exposure control method from Oct 2015 to Dec 2015. Next, we conducted One-way ANOVA test on DLPs by phase and total DLPs of group A, B and C using SPSS and performed independent T-test on group A-B, group A-C, and group B-C respectively. Finally, we also did quantitative analysis by image assessment.

Result : One-way Anova test on total DLP values of group A, B, and C using SPSS program showed that average DLP value of group C is significant (P<0.05) in 95% confidence interval. Besides, group B in A-B, group C in A-C. and group C in B-C respectively showed their average total DLP is significant (P<0.05) in 95% confidence interval and so we confirmed dose reduction in group C.

Conclusion : For the reduction of patient dose in Urography CT, we studied the usefulness of relationship between BMI and patient dose to apply the optimal condition to clinical diagnostic.
KEYWORD
BMI, DLP, Unography CT, Dose Reduction, Care kV, Care Dose