KMID : 1134820130420020244
|
|
Journal of the Korean Society of Food Science and Nutrition 2013 Volume.42 No. 2 p.244 ~ p.248
|
|
Confirmation of Schizandrin as a Marker Compound in Jangsu Omija Powder
|
|
Kim Yun-Jeong
Ha Na Han Song-Hee Jeon Ji-Young Hwang Min-ho Lee Sun-Young Chae Soo-Wan Kim Min-Gul
|
|
Abstract
|
|
|
We studied the development of schizandrin as a marker compound in Jangsu Omija. Schizandrin was validated for its LOD (limit of detection), LOQ (limit of quantitation), precision, accuracy, and recovery by HPLC relative to Omija powder. It showed a high linearity in the calibration curve with a coefficient of correlation (R2) of 0.9990. The LOD and LOQ were 0.2 ¥ìg/mL and 0.5 ¥ìg/mL, respectively. The intra- and inter-day precision of schizandrin calibration standards was 0.06¢¦0.66% and 0.13¢¦1.19%, respectively, and the intra- and inter-day accuracy of schizandrin was 98.35¢¦103.30% and 98.35¢¦103.00%, respectively. Overall, schizandrin was validated through our analytical methods as a marker compound in Jangsu Omija.
|
|
KEYWORD
|
|
schizandrin, method development, Omija, HPLC, validation
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|